dc.identifier.citationBadla, C. & Wewers, F. 2020. Optimization of x-ray fluorescence calibration through the introduction of synthetic standards for the determination of mineral sands oxides. South African Journal of Chemistry, 73: 92-102. [https://doi.org/10.17159/0379-4350/2020/v73a14]en_US
dc.identifier.issn0379-4350-
dc.identifier.urihttp://hdl.handle.net/11189/8660-
dc.description.abstractAn XRF spectrometer traditionally requires its methods to be calibrated using certified reference materials (CRMs). This means choosing from a limited selection of CRMs, which are costly and not widely available. This article examines the application of synthetic oxide standards made from commercially available high purity compounds for the calibration of an XRF spectrometer. Enabling it to analyze for major, minor and trace elements (Mg, Al, Si, P, Hf, Ca, Ti, Zr, Fe, Th, U) in heavy mineral sands processing. Calibration standards were synthesized by mixing pure compounds and mineral sands reference materials, to mimic matrices similar to that of the routine samples and cover the required analytical range. The aliquots were mixed in % fractions and fused with a 1:9 sample/borate dilution to make glass beads. The optimized calibration was validated as per SANAS TR 26-02 criteria for linearity, working range, limit of detection (LOD) and quantification (LOQ), precision and accuracy. The correlation coefficients (r2 ) of the resulting calibration curves were >0.999, showing excellent linearity. The working range for all the calibration lines was satisfactory. The LOD and LOQ values for all the lines were low enough to prevent extrapolation below the lowest calibration point. The precision of the calibration was sufficiently high, and the accuracy was of adequate quality with z(MAD) <2. These observations support the successful synthesis and use of a well-selected set of synthetic standards. Additionally, it has been successfully utilized to analyze a variety of heavy mineral sands samples.en_US
dc.language.isoenen_US
dc.publisherSouth African Chemical Instituteen_US
dc.relation.ispartofSouth African Journal of Chemistryen_US
dc.subjectMineral sanden_US
dc.subjectXRFen_US
dc.subjectCRMsen_US
dc.subjectcalibrationen_US
dc.subjectsynthetic standardsen_US
dc.subjectvalidationen_US
dc.titleOptimization of x-ray fluorescence calibration through the introduction of synthetic standards for the determination of mineral sands oxidesen_US
dc.identifier.doihttps://doi.org/10.17159/0379-4350/2020/v73a14-
dc.typeArticleen_US
Appears in Collections:Appsc - Journal Articles (DHET subsidised)
Files in This Item:
File Description SizeFormat 
Optimization_X-ray_Fluorescence.pdf983.17 kBAdobe PDFView/Open
Show simple item record

Page view(s)

124
Last Week
0
Last month
4
checked on Sep 2, 2026

Download(s)

184
checked on Sep 2, 2026

Google ScholarTM

Check

Altmetric


Items in Digital Knowledge are protected by copyright, with all rights reserved, unless otherwise indicated.